| Sperm Preparation
|
Percoll Separation Procedure Place
the Percoll column with the semen layerd on top in the centrifuge.
You need to centrifuge at 700 g for 15
minutes. Use the
following formula to determine the rpm needed to run the centrifuge. [Relative
Centrifuge Force (G) = (1.119 X 10-5) X rpm2 X
radius(cm)].
You must come up with
the correct speed before progressing furthe! |