1. The hemocytometer is a device for counting cells or particles. As you can see below it is composed of a thick piece of glass with 2 rails on each side. The rails are designed to hold a coverslip 0.1 mm above a mirrored surface in the center. This surface has a grid etched into it. Be sure to clean the hemocytometer and the coverslip before beginning using a little water or alcohol followed by drying with a kimwipe. |
|||||||||||||||||||||||||||||||||||||||
![]() |
|||||||||||||||||||||||||||||||||||||||
2. Prepare an incubation chamber (large petri dish with supports for hemocytometer and wet paper in the bottom). 3. Dilute sperm with water. The objective is to kill sperm and to dilute them sufficiently so that you will count 100 - 400 sperm cells in the 5 squares as described below. The exact dilution needs to vary with the species and concentration of the ejaculate. A general guide line follows: |
|||||||||||||||||||||||||||||||||||||||
|
|||||||||||||||||||||||||||||||||||||||
4. Place a coverslip on the hemocytometer after wetting supports (rails) with saliva or water. This helps to hold down the coverslip while loading the sperm. 5. Place 10-15 µl of the diluted sperm under the cover slip on each side of the hemocytometer as shown in the figure 1a above. 6. Carefully place the hemocytometer in the prewetted chamber, close the lid and wait 5 minutes. 7. Remove the hemocytometer without tilting it, dry the bottom if needed and place on a microscope. View the hemocytometer with a 40 X objective. 8. If you have used a phase type hemocytometer (the thin ones) then be sure the phase rings are in and the condensor is all the way up. You should be able to focus on the sperm but be careful not to break the coverslip. If you used one of the thicker hemocytometers, pull the phase rings out and close down the apeture on the condensor so that you can see the sperm. You may also need to raise or lower the condensor so you can see the sperm. |
|||||||||||||||||||||||||||||||||||||||
9. Count the sperm in 5 of the squares as shown in the following images. Be sure you count 5 squares on each side of the hemocytometer and then use the average for your calculations.
|
|||||||||||||||||||||||||||||||||||||||
10. The hemocytometer is 0.1 mm deep and the 25 large squares represent an area of 1 square mm. The volume above the 25 squares shown is 0.1 µl. As you are only counting 5 squares, you counted the sperm that settled out of 0.02 µl (0.1/5=0.02). Therefore you must multiply your count in 5 squares by 50,000 in order to determine how many sperm would have been in 1.0 ml (1000/0.02=50,000; we usually express sperm concentration in terms of numbers/ml). To get the concentration of the original sperm sample we must however also multiply by the dilution factor. The equation that follows would be used to convert the counts in 5 squares to concentration/ml: Concentration/ml = (Dilution Factor)(Count in 5 squares)(0.05 X 106) By convention, sperm concentration is usually expressed in terms of sperm X 106/ml. |
|||||||||||||||||||||||||||||||||||||||